●In line with requirements of the US NSF/ANSI49 standard for Class II biological safety cabinets to be classified as Type A2, with corresponding UL and NSF certificates.
●Vertical laminar negative pressure model. 70% of the air flow is filtered and recycled, and 30% of the air flow can be discharged into the room or connected to the exhaust system after being filtered.
●High-definition LCD human-machine dialogue interface, touch key operation, real-time display of wind speed, filter operation status, and provide fault alarm.
●The imported intelligent air volume automatic compensation system ensures that the air volume change of the fan is less than 10% when the filter resistance increases by 50%, which improves the safety.
●Each biological safety cabinet has passed the factory test of safety and product performance in accordance with international standards.
●In line with requirements of the US NSF/ANSI49 standard for Class II biological safety cabinets to be classified as Type A2, with corresponding UL and NSF certificates.
●Vertical laminar negative pressure model. 70% of the air flow is filtered and recycled, and 30% of the air flow can be discharged into the room or connected to the exhaust system after being filtered.
●High-definition LCD human-machine dialogue interface, touch key operation, real-time display of wind speed, filter operation status, and provide fault alarm.
●The imported intelligent air volume automatic compensation system ensures that the air volume change of the fan is less than 10% when the filter resistance increases by 50%, which improves the safety.
●Each biological safety cabinet has passed the factory test of safety and product performance in accordance with international standards.
Friendly Control Panel ◆ LED screen, show the air velocity, time, fan station, alarm in real time. | ![]() |
Safe and Reliable, Energy-Efficient ◆ DC fan and LED light make the operation power consumption save 30%. ◆ Intelligent auto-supply system of imported blower ensures the change of air volume less than 10% when the resistance of filter increases 50% and enhance safety. ◆ The alarm system for safe operation height is interlocked with light. At the same time light is Interlocked with sterilizing. ◆ Leak detection in cabinet, no leakage at 500Pa. ◆ UV light with set timer, safer and high efficiency.
Humanized design ◆ 10°sloped front make it comfortable to operator. ◆ Negative pressure and double layers chamber around ensure non-leakage, non-pollution. Demountable circular-arc armrest can alleviate working tiredness and make it easy to transport. ◆ Front window adopts hanging lifting system (Imported from Japan) and more than 5mm thickness security glass with arbitrary positioning, high stability and convenient to sterilization when it closed. ◆ Movable stainless steel worktable 4 liter capacity stainless steel collecting tank with below down valve, easy to collect water, cleaning and disinfection. ◆ Non-obstructive air return mode in the front suction channel. ◆ The adjustable leveling foot is without exposure thread. |
|
HEPA filters ◆ USA HV HEPA Filter Medium, provide 99.995% typical efficiency for particle size over 0.3 microns, and the materials is glass fiber. ◆ A minimum ISO 5 air quality in the work area ◆ Unique Technology of non leakage. ◆ Every HEPA filters will be tested to ensure non leakage. |
|
Model/Item | BSC-1301ⅡA2 | BSC-1302ⅡA2 | |
| Nominal Size | 1.3meter(4') | ||
Filtration Efficiency | HEPA: ≥99.995%, @0.3μm/ULPA:≥99.9995%, @0.12μm | ||
Downflow Velocity | 0.35 m/s(68.88fpm) | ||
Inflow Velocity | 0.54 m/s(106.27fpm) | ||
Noise | ≤67 dB(A) | ||
| Vibration(μm) | ≤5 μm(center of tabletop) | ||
Power Supply | AC115V,1φ,60Hz | AC230V,1φ,50/60Hz | |
| Opening Height(mm) | 250 | ||
Operator Protection | A. Total colony in impaction sampler ≤10CFU./time B. Total colony in slot sampler ≤5CFU./time | ||
Product Protection | Total colony in culture dish ≤5CFU./time | ||
Cross Contamination | Total colony in culture dish ≤2CFU./time | ||
Max Consumption(with spare socket) | 1.5 kVA | / | |
| Rated Power(without spare socket) | 1 kVA | 0.46 kVA | |
Weight | 300 kg | ||
Work Dimension (W1XD1XH1)
| 1270*600*620 mm (50''x23.6''x24.4'') | ||
Overall Dimension (WXDXH)
| 1500*790*2050 mm (59.1''x31.1''x80.7'') | ||
Size and Qty. of Light | LED 24.5 w x ② | ||
Size and Qty. of UV Light | 15 w x ② | ||
Lumin. | ≥750 Lx | ||
| Cabinet Materials | High-grade steel and lacquered in ivory | ||
| Working Area Materials | SS304 fully finished | ||
Air Direction | Top out | ||
Plug Style | USA Type(AC125V/10A) | / | |
Qty. of Spare Socket | USA Type(AC125V/2A) 2 units | / | |
| No. | Name | Note |
| 1 | Electric Socket | Standard, 2 pcs |
| 2 | U.V.Light | Optional |
| 3 | Canopy Connecter | Optional |
| 4 | Gas Valve | Optional |
| 5 | Vacuum Valve | Optional |
| 6 | Water Valve | Optional |
| 7 | I.V.Pole | Optional |