With an extensive track record of safety, reliability and performance, HFsafe LC cabinets make ideal investments for a wide range of applications including work with infectious agents that require Biosafety Level 1, 2 or 3 containment.
On-compromising safety

Enhanced comfort and convenience

| UV Decontamination | |
|
|
Easy to Clean

How to set a new bechmark for the whole industry?

Friendly Communication

Comprehensive Performance Testing
Every HFsafe model manufactured by Heal Force is individually tested, documented by serial number and validated with the following test methods.

| Inflow/Downflow velocity | Air particle count tes | UV radiation test |

| Scan test for filters | Vibration test | Illumination test |
KI Discus containment test according to EN12469 (Potassium Iodide)
![]() | One of the few companies in the world equipped to perform the KI Discus test for validating the operator/personnel protection capabilities Excellent correlation with the microbiological test method for operator protection, and is useful for validating the actual containment performance of the cabinet on-site. The KI-Discus takes only 45 minutes as opposed to 2 days for microbiological testing. Thus, Each HFsafe model is factory tested on a sampling basis using the KI-Discus method for operator safety. |
| Model | HFsafe-1200LC |
| Nominal Size | 1.2 meters(4,) |
| External Dimensions with Base Stand (WxDxH) | 1323x771x2255mm (52.1"x30.4"x88.8") |
| Internal Work Area, Dimensions(WxDxH) | 1200x600x678mm (47.2"x23.6"x26.7") |
| Internal Work Area,Space | 0.72m2(7.8 sq.ft) |
| Average Airflow Velocity * | |
| Inflow | 0.53m/s(104.3fpm) |
| Downflow | 0.35m/s(68.9fpm) |
| Airflow Volume | |
| Inflow | 477m3/h(281cfm) |
| Downflow | 866m3/h(510cfm) |
| Exhaust | 477m3/h(281cfm) |
| ULPA Filter Typical Efficiency | |
| Downflow | Filters provide 99.9995% typical efficiency for particle size of 0.1 to 0.2 microns |
| Exhaust | Filters provide 99.9995% typical efficiency for particle size of 0.1 to 0.2 microns |
| Biosafety Protection Test | |
| Personnel Protection lest | Kl-Discus containment and microbiological testing is performed |
| Product Protection Test 1~8x106(three in succession) | W5CFU |
| Cross-contamination Test 1~8x106(three in succession) | <2CFU |
| Sound Emission (Typical) | |
| NSF/ANSI 49 | <62dBA |
| EN 12469 | <59dBA |
| Fluorescent Light Intensity | 800-1200 Lux (74-112 foot candles) |
| Excellent light distribution | Yes |
| RMS | <2.3 um |
| Cabinet Construction | |
| Main Body | 1.2mm(0.05") steel with white oven-baked epoxy-polyester |
| Work Zone | 1.5mm(0.06") stainless steel, type 304 |
| Side Walls | 1.5mm(0.06") stainless steel, type 304 |
| Electrical sliding windows Option | Yes |
| Window material | Hardened/laminated safey glass |
| Electrical | |
| Cabinet Full Load Amp(FLA) | 2A |
| Fuses | 10A |
| Cabinet Nominal Power | 452W |
| Optional Outlets FLA | 5A |
| Total Cabinet FLA | 7A |
| Power Supply* | |
| 220V/50HZ | Yes |
| 220V/60HZ | Yes |
| 110V/60HZ | Yes |
| Net Weight | |
| Manual Type | 225kg(496lbs) |
| Shipping Weight | |
| Manual Type | 295kg(650lbs) |
| Shipping Dimensions MaximumfW^D^H) | 1425x945x1717mm (56.1"x37.2"x67.3") |
| Shipping Volume, Maximum | 2.30m3(81.2cu.ft.) |

Copy product links
Long by picture save/share
enquiry form:
Inquiry Content:
You have no items to require

enquiry form:
Inquiry Content:
You have no items to require